SCHIPPERS-MEDIZINTECHNIK
H i g h - T e c h i n U r o l o g i c a
l I n s t r u m e n t a t i o n |
SCHIPPERS-MEDIZINTECHNIK
Boschstr. 6
D- 82178 Puchheim / Munich / Germany
Tel.: +49 89 802245 - Fax: +49 89 8003064
Email: info@schippers-med.com |
|
|
|
|
Modulary physiological measuring system
UROMASTER
Technical
data:
|
Flow
transducer: (option) |
| Flow volume: |
|
|
|
|
| Measuring range: |
|
up to
|
2000 ml |
|
| Accuracy: |
|
up to |
50ml |
+/-
1ml |
|
50ml |
up to |
100ml |
+/-3ml |
|
100ml
|
up to |
2000ml |
+/-3% |
| Linearity: |
<
0,05% |
|
|
|
|
|
|
|
|
| Flow rate: |
|
|
|
|
| Measuring range: |
0 |
up
to |
12.5ml/s |
|
|
0 |
up
to |
25ml/s |
|
|
0 |
up
to |
50ml/s |
|
| Accuracy: |
+/-3% |
|
|
|
| A/D-Converter: |
12bit |
|
|
|
Volume transducer (Infusion): |
| Measuring range: |
|
up
to |
2000 ml |
|
|
Scale: |
0 |
up to |
1500ml |
|
|
0 |
up
to |
750ml |
|
|
0 |
up to |
375ml |
|
| Accuracy: |
|
up to |
50ml
|
+/- 1ml |
|
50ml
|
up to |
100ml |
+/-3ml |
|
100ml
|
up to |
2000ml |
+/-3% |
| Linearity: |
< 0,05% |
|
|
|
| A/D-Converter: |
12bit |
|
|
|
Pressure transducer: |
| Measuring range: |
-40 cmH2O |
up to |
400cm H2O |
|
|
Scale: |
0 |
up to |
200cm H2O |
|
|
0 |
up
to |
100cm H2O |
|
|
0 |
up to |
50cm H2O |
|
| Frequency range: |
DC up to 20 Hz |
|
|
|
| Accuracy:
|
|
up
to |
100cm H2O |
+/-2cm H2O |
|
100cmH2O |
up
to |
400cm
H2O |
+/-2% |
| A/D-Converter: |
10bit |
|
|
|
EMG: (option) |
| Measuring range: |
|
up to |
2500µV |
|
|
Scale: |
0 |
up tos |
25µV |
|
|
0 |
up to |
250µV |
|
|
0 |
up to |
2500µV |
|
| Average time: |
0.1sec. |
|
|
|
| Filter: |
100 Hz |
bis |
10 kHz |
|
| A/D-Converter: |
10bit |
|
|
|
Puller device: |
| constant pulling
speed |
| Pulling range: |
280mm |
|
|
|
| Pulling speed: |
0,2
0,5 1 2 5mm/sec. |
| Accuracy: |
+/-2% |
Roller pump: |
| constant
fillingspeed: |
5 ml/min up to 100 ml/min |
| Perfusions quantaty: |
5 and 10ml/min
(Profilometry) |
| Filling speed: |
5,
10, 20,
50 and 75ml/min |
| Accuracy: |
+/-5% |
|
|
|
|
|
|
|
|
|
|
 |